Basic Info.
Product Description
AM-II(AA) Pulse Oximeter
The pulse oximeter is a non-invasive, handheld patient pulse oximeter. It operates only on Common 1.5V AA alkaline or rechargeable battery power. It is compact, small, light, and easy for learning and handling. It is suitable for pulse oximetering adult, pediatric, infant and neonate patients. It is widely used in the hospital's operation room, ICU, clinic section office, out-patient department, sickroom, emergency treatment, and the recovery and health care organizations, or in the family nursing and in the process of transporting patients. It can also apply to animals selectively.
Parameters measured by the pulse oximeter include: arterial oxygen saturation (SpO2), pulse rate (PR), pulse strength. The pulse oximeter measures these parameters through a SpO2 sensor and displays them on the LED screen after certain further processing.
The pulse oximeter is operated and controlled by the buttons on the front panel. It adopts LED in displaying measurements and in supplementary status indication.
The pulse oximeter has the following functions:
- Measuring: Display of SpO2, PR and pulse strength.
- Prompting: Battery capacity.
- Alarm: Judgment of exceeding the limits, Audio and visual alarm.
- Power saving: Automatic shutdown when no pulse oximetering.
Standard Packing List:
- Pulse Oximeter main unit 1pc
- Adult finger SpO2 sensor 1pc
- The operator's manual 1pc
Selectable Accessory:
- pediatric soft tip SpO2 sensor
- Neonate/infant wrap SpO2 sensor
- Ear clip SpO2 sensor
Specifications
General
Pulse oximetering Parameters:SpO2,PR,Pulse strength
Signal sockets:SpO2 socket
Display screen
Type:Red LED
Display area:56mm×36mm
Size:120mm×63mm×32mm
Weight:125g (not include sensor and battery)
Features
Audio and visual alarm:The speaker sounds the alarm and the parameter's measurement flashes on the screen.
Prompting: Indicator of battery capacity.
Power saving: The pulse oximeter will automatic shut down when no pulse oximetering for 3 minutes.
Electrical specifications
Working Voltage:D.C. 2.8V~D.C. 5.4V.
Battery
Type:Three common 1.5V AA alkaline or rechargeable batteries
Run Time:7-hour continuous operation with a new battery (environment temperature is 25ºC)
Environment
Temperature
Operation:0~45ºC
Transportation and storage:-20~60ºC
Humidity
Operation:15%~95 %( noncondensing)
Transportation and storage:10%~95%(noncondensing)
Altitude
Operation:86KPa~106KPa
Transportation and storage:50KPa~106KPa
Parameter Specifications
SpO2
Patient:Adult, pediatric, infant, neonate, animal
Range:35~99%
Resolution:1%
Accuracy:2%(70%~99%),unspecified(0%~69%)
PR
Range:30~250bpm
Resolution:1bpm
Accuracy:2bpm